Date |
Jul Date |
Time |
Battery (V) |
T (ºC) |
Power (W) |
Disk (GB) |
Usage (%) |
M1 (V) |
M2 (V) |
M3 (V) |
M1f (V) |
M2f (V) |
M3f (V) |
Task |
Comments |
Techs |
2008-01-17 | 2008/017 | 08:58:00 | 11.1 | 10.0 | 0.0 | 19.53 | 0.1 | 3.78 | -3.77 | -1.85 | -0.06 | -0.01 | 0.19 | Change of disk and centering the masses. | The bubble was not centered | DN, GM |
2008-03-03 | 2008/063 | 08:45:00 | 12.5 | 15.8 | 0.0 | 19.53 | 13.3 | -1.15 | -0.48 | 1.61 | -0.04 | 0.08 | 0.01 | Centering the masses | Direct battery 14 VDC connected with the solar panels. Those panels are connected with the batteries through a charge regulator | DN, GM |
2008-08-13 | 2008/226 | 09:17:00 | 12.1 | 36.6 | 0.0 | 19.53 | 46.4 | -3.89 | 3.69 | 1.87 | 0.02 | -0.04 | 0.16 | Change of disk. Centering the masses | Bad data from day 198. | DS, GM |
2009-02-26 | 2009/057 | 11:41:00 | 12.8 | 19.4 | 0.0 | 19.53 | 56.4 | -3.66 | -3.71 | 0.03 | 0.09 | 1.22 | 0.04 | Masses' centering. Change of disk | Difficulties found while tunning M2. Replaced the jerrycan | GM, JP |
2009-04-27 | 2009/117 | 11:41:00 | 12.8 | 19.4 | 0.0 | 19.53 | 56.4 | -0.37 | 1.23 | -0.31 | 0.00 | 0.00 | 0.00 | Dismantling. Data unavailable. | M2 is unable to be tunned. Brought to the lab to verify the status of Taurus, wire and Trillium. | DN, JP |
2009-06-03 | 2009/154 | 11:03:00 | 13.7 | 29.6 | 0.0 | 19.56 | 0.7 | -0.37 | 1.23 | -0.31 | 0.00 | 0.00 | 0.00 | Install | Sensor wire fixed. CF used instead of HD | DN, JP |
2009-07-07 | 2009/188 | 00:00:00 | 13.7 | 29.6 | 0.0 | - | - | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | Dismantle | CF removed | DN, GM |